62
INSP/TESTS No. CONDITIONS
AND
REQUIREMENTS
INST/LCC# RESULTS PASS
115 V 230 V
DETERGENT CUP
Audible
HEATER 9.3 – 11.5 Amp 7.8 – 9.6 Amp
Ammeter __________
Return to TAP – Allow water to fill washer until it turns off. Press and
hold RUN – Water must turn off before it overflows.
Visual
DRAIN / COOL – Allow all water to drain
Visual
DRY
BLOWER
Audible
LIQ DETERG PUMP (No test)
LIGHT (Window models only)
Visual
FAN
Aubile
Run Test
1
Fill Detergent Cup with Detergent. Close Door
2
Scroll
through
Programs
Confirm Display Information
Visual
3
115V Models – Select Glass Plus 230V Models- Select Glass Plus
Wash 2 Temp 82°C Wash 2 Temp 93°C
Wash 2 Time 10 min Wash 2 Time 10 min
Steam Yes Steam Yes
No. Rinses 3 No. Rinses 3
No. Pure Rinses 1 No. Pure Rinses 1
Final Rinse Temp 60°C Final Rinse Temp 60°C
Dry Temp 60°C Dry Temp 60°C
Dry Time 20 minutes Dry Time 20 minutes
Start Data Acquisition Computer – Identify file with washer serial
number and date
Press RUN
Confirm operation of washer, temperatures and times
Acceptance Criteria:
Visual-
RS232
Wash 2 and Final Rinse Temp +/- 3°C
Dry Temp +/- 8°C once reached
Wash 1, Steam, Rinses 1 & 2, Dry Time +/- 1 minute
Wash 2, Final Rinse Time after reaching SP +/- 1 minute
4
All Window Models
Observe Wash Action
FlaskScrubbers – Water Hits Top of Tank from spindles
Visual
SteamScrubbers – Lower Wash – Arm Rotates at least 45 RPM
Stop Watch
5
Run rinse only to wet surfaces.
Visual
After wash pump starts, place a wet towel across the overflow
contacts. Wash pump should stop and the washer should drain.
Display should show ALARM OVERFLOW.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com