Chapter 5: Using The Ductless Enclosure
Product Service 1-800-522-7658
35
Filter Setup Display Options
This menu selection determines which filters are loaded and displayed in
the lower and upper positions. One of seven filter types must be selected
for each filter position.
The filter check time and filter final life must be selected next for each
filter position. Consult Chapter 3, Appendix E, the Labconco Specialist,
and the Paramount Chemical Guide for proper program times.
Restoring Software Default Settings
To restore the default settings to their original configuration:
1.
Access the Menu, and select “SETUP.” Press “ENTER.”
2.
In the Setup submenu, select “Restore Defaults.” Press “ENTER.”
Table of Default Settings
Blower: Smart-Flow
enabled
Units of Measure:
fpm
Security Lock:
OFF, Non-secure
Timers:
Blower and Light Timers disabled.
Filter Timers:
Enabled; Interval set at 100 hours, Final set at
500 hours.
Calibration
This password-protected selection is for qualified certifiers to adjust the
blower speed, face velocity, and the low airflow alarm setpoint. Certifier
access code is set at 123.
Diagnostic Test Operation
When the “Diagnostic Test” menu is selected, the LCD allows “Keypad
Test” to help troubleshoot any malfunction. The “Pressure Sensor”
displays pressure values to help troubleshoot an airflow malfunction.
The “Organic Sensor” displays chemical breakthrough (millivolt
readings) and alarm setpoints to troubleshoot an organic sensor
malfunction.