Chapter 7: Troubleshooting
Product Service 1-800-522-7658
29
PROBLEM
CAUSE
CORRECTIVE ACTION
Frequent oil change needed in
pump
Secondary Trap insert is spent
Change insert often.
Cold Trap is not emptied after
each run and dried
Empty the traps (glass or stainless steel) after
each run and replace.
Vacuum too strong for chemical
Use a Secondary Trap insert and diaphragm
pump.
Unit starts and shuts off
Latch optical sensor improperly
calibrated
Unplug power cord. Wait 10 seconds. Plug in
power cord.