Chapter 3: Getting Started
Product Service 1-800-522-7658
8
Setting Up Your CentriVap
After you verify receipt of the proper components, move your CentriVap to the
level and sturdy location where you want to install it. There is no need to secure
the CentriVap to the work surface. Then, follow the steps listed below.
Emergency Access Into the Chamber
The CentriVap is designed to prevent access to the chamber in the event of a
power disruption. If it is necessary to open the lid when there is no electrical
power connected to the CentriVap, insert a small screwdriver or similar
instrument into the small round hole on the left side of the case near the top
behind the control panel. This will unlock the lid latch mechanism. While
holding the screwdriver in place, raise the lid with the other hand.
CAUTION: Never attempt to defeat the latch or open the lid while the
CentriVap is running. Personnel injury can result from moving parts and
chemicals.