Chapter 4: Using Your CentriVap Mobile System
Product Service 1-800-522-7658
19
If during a run the “STOP” button is pressed, the timers remember the time at
which stop occurred. If “RUN” is then pressed, the timers continue to count
down from the time at which they were stopped. If you are running program 1,
program 2 or program 3, pressing “PROG 1,” “PROG 2” or “PROG 3” resets the
timers to the original set point time and the CentriVap starts a new run.
To reset the timers to the original set point time when running programs other
than program 1, program 2 or program 3, press the start button and hold it for five
seconds. The display will indicate that the timer is reset.
Operating the CentriVap
CAUTION: To avoid personnel injury; Do not operate the CentriVap if the
lid is scratched or nicked, or shows signs of damage. A damaged lid could fail
under vacuum.
While the CentriVap Concentrator is operating, do not lean on the lid, do not
stand near it longer than necessary and do not place hazardous materials within 12
inches.
1.
Turn “ON” the CentriVap power switch.
2.
Press the Cold Trap “ON” switch. The “Y” in the display under the cold trap
will flash. When the trap reaches operating temperature the “Y” will
illuminate steady.
3.
Select a program or set the set point parameters.
4.
Preheat the chamber if desired.
5.
Turn on Heat Boost if the CentriVap is so equipped and additional heat is
desired.
6.
Place samples in vials. Normally the vials should be no more than half full.
Place vials in a rotor.
7.
Load the rotor with samples into the chamber.
8.
Close the lid.