Chapter 2:
Prerequisites
Before you install your CentriVap, you need to prepare your site for installation.
Carefully examine the location where you intend to install your CentriVap. You
must be certain that the area is level and of solid construction. In addition, an
exhaust means must be provided. An electrical source must be located near the
installation site.
Carefully read this chapter to learn:
The electrical supply requirements.
The exhaust requirements.
Refer to
Appendix C: CentriVap Specifications
for complete CentriVap electrical
and environmental conditions, specifications and requirements.
Electrical Requirements
The CentriVap Concentrator requires a dedicated grounded electrical outlet. This
outlet requires a 15 Amp circuit breaker or fuse for models rated at 115V (60 Hz).
An 8 Amp circuit breaker or fuse is required for models rated at 230V (50/60 Hz).
Various models are available for use in different locations throughout the world.
Each has a unique plug suitable for the specific location. See CentriVap
Components in Chapter 3 for detail plug specifications. If the plug supplied with
the CentriVap Does not match with the available receptacle, remove this plug and
replace it with an approved plug of the suitable style.
It is recommended that an emergency switch for disconnecting the mains in the
case of a malfunction be located remote from the CentriVap, preferably outside
the room in which the CentriVap is housed, or adjacent to the exit from that room.
Summary of Contents for CentriVap DNA Systems
Page 1: ...CentriVap DNA Systems Models 7970010 7970011 7970030 7970035 7970037...
Page 32: ...Appendix A CentriVap Components Product Service 1 800 522 7658 28...
Page 33: ...Product Service 1 800 522 7658 29 Appendix B CentriVap Dimensions...
Page 35: ...Appendix C CentriVap Specifications Product Service 1 800 522 7658 31...