Product Service 1-800-522-7658
61
A
PPENDIX
C
B
LOWER
E
NVIRONMENTAL
C
ONDITIONS,
W
IRING
D
ETAIL &
W
IRING
D
IAGRAMS
Environmental Conditions
Maximum altitude: 9843 feet (3000 meters).
Ambient temperature range: -30° to 130°F (-34° to 54°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.