109
9. Remove the 3 supply HEPA filter retainer brackets (see Figure 12-15).
10. Lift up on one corner of the supply HEPA filter to break the seal between the filter
gasket and the filter shelf. Once the filter is free, slide it straight out.
Reinstallation Notes:
1. When reinstalling the supply HEPA filter, to lower the plenum onto the supply
HEPA, turn the 3 plenum lift bolts counter-clockwise. Turn each bolt part way,
working in a pattern amongst the 3 bolts to lower the plenum evenly.
2. When downward pressure begins to be applied onto the supply HEPA (and
upward pressure on the exhaust HEPA when it is reinstalled separately), the
brackets holding the bolts you are tightening will flex slightly. Only tighten until
the filters are secure and the filter gasket has 50% compression. Do not
overtighten the bolts.
Figure 12-15
Filter Retainer Bracket
Summary of Contents for Logic+ A2
Page 149: ...149 208 240V...
Page 151: ...151 Blower only will not start...
Page 152: ...152 Lights only will not illuminate...
Page 153: ...153 UV Light will not illuminate...
Page 154: ...154 Airflow Alert activating...
Page 155: ...155 Filter Life Gauge not at 100 when new...
Page 156: ...156 Contamination in the work area...
Page 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Page 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...