Calibration
After airflow sensor installation, you MUST calibrate the new sensor. Follow these
steps to calibrate the new sensor.
Note:
To properly calibrate the airflow sensor, a qualified certifier must verify the inflow
and downflow values.
1. From the Home Screen, press
[MENU]
.
2. Press
[DOWN]
until the
Tools
option is highlighted.
3. Press
[OK/MUTE]
to enter the menu screen.
4. Press the
[DOWN]
button until the
Password
option is
highlighted, then press
[OK/MUTE]
to select the
certifier’s submenus.
5. You will see a
Warning Screen
alerting you that you
are about to alter the BSC’s settings; Press
[OK/MUTE]
to input the password.
6. When requested for the password press
[LIGHT]
,
[UV LIGHT]
,
[TIMER]
,
[OUTLET]
then
[OK/MUTE]
.