Safety Precautions for this Product
Electrical outlets in the cabinet are restricted to 5 amps (100-115v) or 3 amps
(230v) maximum current.
Prises électriques dans l'armoire sont limitées à 5 (100-115v) o 3 (230v) courant
maximum ampères
.
Do not use any detachable power cord that is not adequately rated for the unit.
Ne pas utliser un fil électrique amovible qui n’est pas du tension nominale de
l’appareil.
The biosafety cabinet should be certified by a certification technician before its
initial use. The cabinet should be recertified whenever it is relocated, serviced or
at least annually thereafter. Filter integrity and airflow performance should be
verified before using the cabinet.
Some internal components of the biosafety cabinet may become contaminated
during operation of the unit. Only experienced personnel competent in
decontamination procedures should decontaminate the cabinet before servicing
these components. If you have any questions regarding certification agencies, or
need assistance in locating one, contact Labconco’s Product Service Department
at 800-821-5525 or 816-333-8811.
DO NOT load more than 150 lbs. (68 kg) in the work area. Exceeding this limit
may damage the work surface and its supports. Excessive weight in the cabinet
may increase the risk of it overturning, or failure of hydraulic lift stands, resulting
in the cabinet and stand overturning. If your application requires loading more
than 150 lbs. (68 kg)
, contact Labconco’s Product Service Department at 800-
821-5525 or 816-333-8811 for assistance.
Avoid the use of flammable gases or solvents in the biosafety cabinet. Care must
be taken to ensure against the concentration of flammable or explosive gases or
vapors. An open flame should NOT be used in the biosafety cabinet. Open
flames will disrupt airflow patterns, burn the HEPA filter and/or damage the filter’s
adhesive. Gases under high pressure should not be used in the biosafety
cabinet, as they may disrupt its airflow patterns.
HEPA filters only remove particulate matter. Operations generating volatile toxic
chemicals or radionuclides must be evaluated carefully.
Summary of Contents for Logic+ A2
Page 149: ...149 208 240V...
Page 151: ...151 Blower only will not start...
Page 152: ...152 Lights only will not illuminate...
Page 153: ...153 UV Light will not illuminate...
Page 154: ...154 Airflow Alert activating...
Page 155: ...155 Filter Life Gauge not at 100 when new...
Page 156: ...156 Contamination in the work area...
Page 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Page 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...