193
9. Connect user-supplied wires to the appropriate contacts as desired.
Use 22-24
AWG wires ONLY.
Strip insulation back from wire end 0.25 inches (6 mm).
10.
Using a small jeweler’s flat blade screwdriver or similar instrument, push the
screwdriver into the release opening (see Figure 22-6), and rotate the
screwdriver up (away from the wiring connection) slightly. This will open the
clamp. Slide the stripped wire end into the appropriate slot on the connector,
then remove the screwdriver. To release a wire, reinsert the screwdriver, rotate it
up (away from the wire), and pull the wire straight out from the connector. See
Figure 22-6 for reference.
The relay contact closes when the condition is present on the biosafety cabinet. See
the pinout configuration below in Table 22-1 to properly connect the wires.
Figure 22-6
Release Opening
Stripped wire slot
on connector
Position
Contact Label
Condition/Description
1
GND
Common used with +12 VDC
2
RELAY 1
Closes upon Blower start (remains
closed while Blower at Normal Speed)
3
RELAY 1
4
RELAY 2
Closes upon any Alarm Condition
5
RELAY 2
6
RELAY 3
Closes upon Blower at Reduced
Speed (Night Setback
– sash closed)
7
RELAY 3
8
FUSED +12 VDC
+12 VDC supplied (Fused @ 0.5 A)
Table 22-1
Summary of Contents for Logic+ A2
Page 149: ...149 208 240V...
Page 151: ...151 Blower only will not start...
Page 152: ...152 Lights only will not illuminate...
Page 153: ...153 UV Light will not illuminate...
Page 154: ...154 Airflow Alert activating...
Page 155: ...155 Filter Life Gauge not at 100 when new...
Page 156: ...156 Contamination in the work area...
Page 200: ...200 Figure 23 4 1 2 4 3 5 7 6 8 8...
Page 201: ...201 END OF 3849920 Figure 23 5 9 10 11 12 13...