Type C1 Models
1:
Unlike Type B biosafety cabinets, the recommended vacuum will remain constant
throughout the life of the exhaust HEPA filter. Duct vacuums below 0.05 inches H
2
O
(12 Pa) or above 0.5 inches H
2
O (125 Pa) may result in erratic operation and throw an
alarm condition from the product.
2:
WC = Inches of Water Column, typically expressed in units of
inches H
2
O
.
Axiom Type C1 Model
Airflow Volume
Concurrent
Balance Value
Recommended
Duct Vacuum
1
ft
3
/min
m
3
/hr
ft
3
/min
m
3
/hr
WC
2
Pa
4-foot, 8" Sash
323
549
387
658
0.30
75
4-foot, 10" Sash
400
680
480
816
0.30
75
6-foot, 8" Sash
463
787
556
945
0.30
75
6-foot, 10" Sash
570
968
684
1162
0.30
75
Table 7-4