38
6. On a Type A2 or C1 only, if field installing, or removing the Airflow Sensor,
change the selection accordingly, press
[OK/MUTE]
.
7. On all biosafety cabinet types, if field installing, or removing the UV Light, change
the selection accordingly (screen shown above), press
[OK/MUTE]
.
8. On Type A2 models only, if field installing, or removing the Canopy connection,
change the selection accordingly, press
[OK/MUTE]
.
Note:
When a Type C is selected previously, the
Canopy
Screen will not appear, but
the software will default the selection correctly based on whether A-mode (No Canopy)
or B-mode (Canopy) was previously selected.
9. The
Sash Height
selection will display next, and should
NOT
be changed, unless
the model is a C1, and the user has decided to switch between 8-inch and 10-
inch sash operation. Press
[OK/MUTE]
.
If the biosafety cabinet is a Type A2 or B2, this is the final screen selection. If a C1,
proceed to step #10.
AIRFLOW SENSOR
AIRFLOW SENSOR
[OK/MUTE]
UV LIGHT
UV LIGHT
CANOPY
CANOPY
SASH HEIGHT
8 INCH
[OK/MUTE]