160
Canopy (optional)
1. From the Home Screen, press
[MENU]
.
2. Press
[DOWN]
to select the
Tools
option.
3. Press
[OK/MUTE]
to enter the submenu.
4. Press
[DOWN]
until the
Diagnostics
option is
highlighted, then
[OK/MUTE]
to enter the
submenu.
5. Press
[DOWN]
until the
Canopy
option is
highlighted, then
[OK/MUTE]
to enter Canopy
diagnostic.
NOTE:
Type B2 cabinets will not display
Canopy
, because they cannot have a canopy.
6. The display shows the status of the Canopy Inlet
Valve. During normal operation, it will be
OPEN
;
during a canopy alarm, it will be
CLOSED
.
NOTE
: The Canopy Inlet Valve position is
OPPOSITE the magnetic switch condition used
to detect the Inlet Valve position!
– WHEN THE
INLET VALVE IS OPEN (NORMAL), THE
MAGNETIC SWITCH IS CLOSED (completes
the electrical circuit).
7. Press
[MENU]
to return to the Diagnostic submenu.