176
Configuration (new install only)
After airflow sensor installation, you MUST reconfigure the biosafety cabinet to enable
the airflow sensor.
1. From the Home Screen, press the
[MENU]
button.
2. Press the
[DOWN]
button until the
Tools
option is
highlighted.
3. Press
[OK/MUTE]
to enter the menu screen.
4. Press the
[DOWN]
button until the
Password
option is
highlighted, then press
[OK/MUTE]
to select the
certifier’s submenus.
5. You will see a
Warning Screen
alerting you that you are
about to enter an area that can
alter the BSC’s settings;
Press
[OK/MUTE]
to input the password.
6. When requested for the password press:
[LIGHT]
,
[UV LIGHT]
,
[TIMER]
,
[OUTLET]
then
[OK/MUTE]
.
7. If the password is entered correctly, you can select
between altering the cabinet’s configuration, or
calibration. Highlight the
Configuration
option, then
press
[OK/MUTE]
to start the configuration procedure.