Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
19
Original instructions
6.
Energy efficient fluorescent lighting
is located behind a laminated safety
glass shield mounted to the top of the hood. The factory-wired instant
start T8 lighting is serviceable from outside the hood cavity. If needed,
more energy efficient, long lasting 50,000 hour LED direct replacement
T8 bulbs can be ordered. See Appendix A.
7.
Low mounted, factory-wired light and blower switches
are ADA
compliant.
8.
All Air Foils
allow air to sweep the work surface for maximum
containment. The Clean-Sweep™ openings create a constant barrier from
contaminants. In addition, should the operator inadvertently block the
airflow entering, the air continues to enter from under the air foil and
through the Clean-Sweep openings. See Appendix A for alternate PVC
Eco-Air Foil as PVC material resists corrosion from mineral acids.
9.
Streamlined corner posts
provide maximum visibility and the flexibility
to add services after installation.
10.
All hoods are factory prepared for up to 8 service fixtures
(4 service
fixtures if Side View Window option is selected).
11.
Duplex electrical receptacles
are mounted on the right and left corner
posts as requested. Receptacles are factory-wired to hood single point
junction box.
12.
Shipped fully assembled
and eliminates the need for costly onsite
assembly.
13.
Accessory Guardian™ Digital Airflow Monitor
or
Guardian Airflow
Monitor
continuously monitors face velocity. An audio/visual alarm
alerts the user to low airflow conditions. The right corner post is factory
prepared to accommodate the Guardian Monitor (sold separately).
14.
Sash Alarm.
Audible alarm sounds if both sashes are raised to alert user.
15.
Frame of epoxy-coated steel and aluminum
is durable and corrosion
resistant.
16.
Exhaust Connection
. The hood features 12" (12.75" OD pipe) exhaust
connections sized to allow for a minimum static pressure loss through the
hood structure while providing a good transport velocity through the
exhaust system.
17.
Spillstopper™ Solid Epoxy Work Surface
is dished to contain spills.
(Work surface is sold separately).
18.
Optional Ceiling Enclosure Kits
are available for a decorative facade
between the hood and the ceiling.
19.
Optional Sash Stops
provide a means of controlling the operating height
of the sash.