Electrical Data
Model #
Requirements
Energy Use
3410000, -01
115 VAC, 60 Hz, 1 Amp
100W
3410020 through -25
230 VAC, 50/60 Hz, 0.5 Amp
100W
Environmental Conditions
Indoor use only.
Maximum altitude: 6562 feet (2000 meters).
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F
(31°C), decreasing linearly to 50% relative humidity at 104°F
(40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient overvoltages according to Installation Categories II
(Overvoltage Categories per IEC 1010). Temporary voltage spikes
on the AC input line that may be as high as 1500V for 115V models
and 2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally
only non-conductive atmospheres are present). Occasionally,
however, a temporary conductivity caused by condensation must
be expected, in accordance with IEC 664.