Chapter 2: Prerequisites
Product Service 1-800-522-7658
3
Exhaust Requirements
As stated earlier, the exhaust discharge of the station, located on the
upper left side on the back of the station should never be obstructed. The
station is designed to have its back surface pushed against a wall, but the
top of the station should have no obstructions within 6 inches of the
exhaust discharge.
Electrical Requirements
The Downdraft Station models have the following electrical requirements:
Model #
Requirements
3410000-01
115 VAC, 60 Hz, 1 Amp
3410020-25
230 VAC, 50/60 Hz, 0.5 Amp
Note: An outlet with a circuit breaker rated at 10 amps for 115 volt
models (5 amps for 230 volt models) should be located as close as
possible to the right rear side of the station. Power cords supplied
with the station are 3 meters (10 feet) long.
Table 2-1