23
Step 1
Carefully remove the shrink-wrap or outer carton and inspect the product for damage
that may have occurred in transit.
All the filtration fans are packaged in boxes,
secured to the top of the filtered hood with bubble wrap and should not be discarded.
If the product is damaged, take pictures of the product and the outer packaging, and
notify the delivery carrier immediately. Retain the entire shipment, including outer
packaging, intact for inspection by the carrier.
Note:
United States Interstate Commerce Commission rules require that claims be filed
with the delivery carrier within fifteen (15) days of delivery.
Do not return goods without the prior authorization of Labconco. Unauthorized returns
will not be accepted.
If the product was damaged in transit, you must file a claim directly with the freight
carrier. Labconco Corporation and its dealers are not responsible for shipping damages.
Do not discard the carton or packing material for the product until all of the components
have been checked, installed and tested.
Step 2
Note:
Leave the filtered hood attached to its shipping skid until it is as close to its final
location as possible. Move the filtered hood using a suitable floor jack, or by placing a
furniture dolly underneath the skid. Do not move the filtered hood by tilting it onto a
hand truck.
A
DETAIL A
B
DETAIL B
REMOVE SCREW
WITH #2 PHILLIPS
(BOTH SIDES)
LIFT AND REMOVE
BOTH SIDE PANELS
USE 7/16" SOCKET
DRILL TO REMOVE
4 SCREWS IN SKID
Figure 4-1
Summary of Contents for Protector Airo
Page 78: ...78 Fans and Lights not working ...
Page 79: ...79 Vertical sash no longer operates smoothly ...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working ...
Page 81: ...81 Contaminates outside the filtered hood ...
Page 82: ...82 Lights operate but fans will not ...
Page 83: ...83 Smart Command Alarms ...
Page 84: ...84 If needed contact Labconco to troubleshoot further ...
Page 94: ...94 Wiring Diagram Figure C 1 ...