96
(the
XXXX
is the serial number of the filtered hood).
3. Password is:
GFH-2018
4.
Open an internet browser (Google Chrome, Mozilla Firefox… etc.) and enter the
following address in the browser menu:
For connecting the filtered hood to a laptop:
1. Connect an Ethernet cable to the laptop from the command module.
2. Configure the Ethernet parameter of computer:
a. Protocol TCP/IP V4
b. IP address: 192.168.0.205
c. Subnet mask: 255.255.255.0
3.
Open an internet browser (Google Chrome, Mozilla Firefox… etc.) and enter
the following address in the address menu:
4. See Figure 6-3 for more details.
This will open a webpage that is the eGuard
index
home screen. The homepage has
three sections called INFO, LOGS and SETTINGS.
INFO
INFO displays the usage time of the filtered hood as well as the
sensor values of the temperature and humidity of the filtered
hood. The main purpose for the info page is the safety state of
the following categories: Face Velocity, Fan, Filtration and
Maintenance. These correspond to icons on the command
module. When the ventilation is off, the dots beside these
categories are gray. Once the ventilation is activated and the
filtered hood is running correctly, the dot turns green as shown
in Figure D-2.
If any error occurs, that displays with a red dot and the drop
down menu will show the alarm code, giving insight to what is
wrong. The icon and alarm will match the command module
’s
alarm, but through eGuard you will have more information of
the specific problem. If more than one error occurs under a
category there will be more than one error code displayed.
For example,
there could be “Anemometer out of order” and
“Low face velocity” under the same icon. This means two
areas must be addressed before the alarm will cease.
Figure D-3
Figure D-2
Summary of Contents for Protector Airo
Page 78: ...78 Fans and Lights not working ...
Page 79: ...79 Vertical sash no longer operates smoothly ...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working ...
Page 81: ...81 Contaminates outside the filtered hood ...
Page 82: ...82 Lights operate but fans will not ...
Page 83: ...83 Smart Command Alarms ...
Page 84: ...84 If needed contact Labconco to troubleshoot further ...
Page 94: ...94 Wiring Diagram Figure C 1 ...