44
information on decreasing the volume of the alarm see
Outlet (Optional)
If optional outlets found on the filtered hood are electrically connected, they will
operate even if the filtration technology is not currently turned on.
Start-up
1. To begin, make sure the filtered hood is plugged in. You will hear the
power up chime when it is ready to operate.
2. Once the filtered hood is ready, press the
[Main ON/OFF]
. This will turn
on both the lights and the fans. You can validate performance by listening
for the fans and seeing the internal lights are on.
3. On your phone or tablet, open up eGuard and double check all technology
is working properly.
Loading Product / Materials
Only load the materials required for the procedure. Do not overload the
filtered hood.
Do not obstruct the front air foil.
Large objects should not be placed close together. Place large objects on
a short stand to elevate them above the work surface. This permits airflow
to sweep under the equipment.
After loading the filtered hood, wait one minute to purge airborne
contaminants from the work area.
Work Techniques
Keep all materials at least 6 inches inside of the sash, and perform all
contaminated operations as far to the rear of the work area as possible.
Segregate clean materials from contaminated materials in the work area.
Avoid using techniques or procedures that disrupt the airflow patterns of
the filtered hood.
Unloading Product / Materials
Objects in contact with contaminated material should be surface
decontaminated before removal from the filtered hood.
All open trays or containers should be covered before being removed from
the filtered hood.
Shutdown
It is recommended there is a thorough decontamination prior to powering
down the fans and filtered hood.
Summary of Contents for Protector Echo
Page 78: ...78 Fans and Lights not working...
Page 79: ...79 Vertical sash no longer operates smoothly...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Page 81: ...81 Contaminates outside the filtered hood...
Page 82: ...82 Lights operate but fans will not...
Page 83: ...83 Smart Command Alarms...
Page 84: ...84 If needed contact Labconco to troubleshoot further...
Page 94: ...94 Wiring Diagram Figure C 1...