Appendix D: eGuard
The filtered hood operating system provides user control with the touch pad command
module and a Wi-Fi-connection interface called eGuard. eGuard connects the filtered
hood to the
user’s phone, laptop or tablet either through an Ethernet cable into the
command module or by connecting the mobile device to the network of the filtered hood
that will be used.
Command Module
Below is a graphic of the command module with callouts for the different buttons and
icons:
Only the fan, light and main power buttons are available when the hood is on and in
use. The mute alarm button appears when an alarm occurs and disappears once
pressed, or after the alarm is resolved. You cannot unmute the alarm chime once
pressed, but the alarm sound will resume after 10 minutes if the error is not resolved.
The other icons are simplified communication of what is happening with the hood. The
network shows only when your device is connected to the hood. The icon being dark
does not mean it lacks the WIFI or Ethernet to connect, only that the connection is not
currently being made. All other icons are displayed red because they are alarms and
disappear when the error is resolved.
Index
: Connecting to your filtered hood with eGuard
For connecting your phone or tablet to the filtered hood with WIFI:
1. Open the network selector on your device.
2. Select the Wi-Fi network:
GFH-XXXXX-XXXX
SMART HALO INDICATOR
(lighted when fans and lights
are on, blinks when filtered
hood is under alarm.)
NETWORK CONNECTION INDICATOR (indicates
when a device is connected to GFH technology
(works both for WIFI and ETHERNET CONNECTION)
MAINTENANCE ALARM
(indicates that maintenances is
needed. More information is
displayed on eGaurd interface).
MUTE ALARM BUTTON
(appears when an alarm is
activated, and mutes the
sound of the alarm.)
Main ON/OFF button
(turn on or off both fans and lights)
Lights ON/OFF button
(turn on or off lights)
Fans ON/OFF button
(turn on or off fans.)
FILTER REPLACEMENT ALARM
(indicates the primary Neutrodine
Unisorb filters must be replaced)
FAN STATUS ALARM
(appears when one or
more fan is out of order)
FACE VELOCITY ALARM INDICATOR
(appears when FACE VELOCITY IS
BELOW RECOMMENDED LIMITS)
Figure D-1
Summary of Contents for Protector Echo
Page 78: ...78 Fans and Lights not working...
Page 79: ...79 Vertical sash no longer operates smoothly...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Page 81: ...81 Contaminates outside the filtered hood...
Page 82: ...82 Lights operate but fans will not...
Page 83: ...83 Smart Command Alarms...
Page 84: ...84 If needed contact Labconco to troubleshoot further...
Page 94: ...94 Wiring Diagram Figure C 1...