Chapter 2: Prerequisites
Product Service 1-800-522-7658
9
Electrical Requirements
Standard duplex electrical receptacles should be nearby for connecting the
FilterMate Portable Exhauster, an airflow monitor, accessory fluorescent
light or other equipment, such as a balance for weighing operations. For your
convenience both the accessory FilterMate and accessory fluorescent light kit
have auxiliary outlet receptacles. It is recommended that the airflow monitor
be connected directly to the auxiliary switched outlet on the rear of the
FilterMate so the airflow monitor is ON when the blower is ON. The
FilterMate auxiliary switched outlet is rated for 8 amps. For other blower
exhaust systems, it is recommended that the airflow monitor be switched by
the same circuit as the blower exhaust. If this is not possible, then the airflow
monitor may be connected to the switched auxiliary outlet on the accessory
fluorescent light kit. The fluorescent light kit auxiliary outlet can be used for
any accessory under 8 amps.
Space Requirements
The dimensions for the different models are shown in Appendix B: Enclosure
Dimensions.