Chapter 5: Using your XPert or XVS Enclosure
Product Service 1-800-522-7658
26
Work Techniques
•
Keep all materials inside the lower air foil, and perform all
contaminated operations as far to the rear of the work area as possible.
•
Segregate all clean and contaminated materials in the work area.
•
Avoid using techniques or procedures that disrupt the airflow patterns
of the enclosure.
Final Purging
•
Upon completion of work, the enclosure should be allowed to operate
for two to three minutes undisturbed, to purge airborne contaminants
from the work area before shutting down the blower.
Unloading Materials and Equipment
•
Objects in contact with contaminated material should be surface
decontaminated before removal from the enclosure.
•
All open trays, weigh vessels or containers should be covered before
being removed from the enclosure.
Shutdown
•
Turn off the exhaust system and accessory light if provided.