Chapter 6: Maintaining your XPert or XVS Enclosure
Product Service 1-800-522-7658
37
8. Now alter the high exhaust volume with the speed control on the
FilterMate or exhaust damper; alternatively, the face velocity may be
changed by lowering the sash. Measure the average face velocity and
enter the high value on the display. The high value must be at least
20 fpm greater than the low value. The high exhaust volume
calibration will take about 5 seconds.
9. Be sure to enter “DONE” after successfully completing the low and
high calibration set points.
10. Once calibration is completed, go to “RUN” and hit “ENTER”. The
value should read close to the high calibration set point.
11. To lower the face velocity to the operating point, simply alter the
exhaust volume with the speed control on the FilterMate or exhaust
damper. Then recheck the face velocity with an anemometer to
confirm the display on the digital airflow monitor.
Airflow Monitor
Installed
Accessory Light
Enclosure
Figure 6-6
Digital 1000 Installed