Product Service 1-800-522-7658
22
QuickChart – Logic+ Type A1 - Metric
Model
30132
30142
30152
30162
Type
A1
A1
A1
A1
Cabinet Size (in meters)
.91
1.22
1.52
1.83
Sash Opening (inches)
305
305
305
305
Starting Serial #
1303_
1303_
1303_
1303_
Downflow Data
Nominal Avg. Downflow (m/s)
.279 +/-.03
.279 +/-.03
.279 +/-.03
.279 +/-.03
grid # of points (rows x columns)
21 (3x7)
24 (3x8)
30(3x10)
36(3x12)
Grid distance from back & sides
152
152
152
152
Distance between rows (mm)
141
141
141
141
Distance between columns (mm)
104
132
137
140
Inflow/Exhaust Data
Nominal Average Inflow (m/s)
.432 +/-.03
.432 +/-.03
.432 +/-.03
.432 +/-.03
Sash Open Area (Sq. m)
.283
.375
.468
.561
Nominal Avg. Exhaust Vol. (m3/s)
.122
.162
.202
.242
Avg. Exhaust Vol. Range (m3/s)
.115-.129
.153-.172
.190-.214
.228-.257
Secondary Inflow Data
3
Sash Opening Template
1
3+
4+
5+
6+
Sensor distance (mm)
2
102
83
83
102
# of Test points
6
8
10
12
Test point location
*
*
*
*
Nominal Avg. Inflow Vel. (m/s)
1.32
1.75
2.17
1.43
Avg. Inflow Vel. Range (m/s)
1.23-1.39
1.64-1.85
2.05-2.31
1.34-1.51
Correction Factor (CF)
.0929
.0929
.0929
.1700
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
.115-.129
.153-.172
.190-.214
.228-.257
Sash Open Area (Sq. m)
.283
.375
.468
.561
Inflow Velocity Range
(AIV / Sash open area)
.432 +/-.03
.432 +/-.03
.432 +/-.03
.432 +/-.03
HEPA Filter Leak Test Data
Air Displacement (m/s)
.283
.378
.472
.566
Laskin Nozzles needed
1
2
2
2
Theoretical aerosol conc. (ug/l)
3
23
34
27
23
Actual aerosol conc. (ug/l)
4
12
18
15
15
Supply HEPA Data
Width x Depth x Height (in.)
5
36x18x3.06
48x18x3.06
60x18x3.06
72x18x3.06
Performance (CFM)
335
445
555
665
Performance (Pressure in. H
2
O)
.62+/-.1”
.5+/-.1”
.5+/-.1”
<.42”
Labconco P/N
3838400
3838401
3838402
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
5
18x18x5.88
26x18x5.88
36x18x5.88
48x18x5.88
Performance (CFM)
266
353
441
529
Performance (Pressure in. H
2
O)
< 0.35”
< 0.28”
< 0.28”
< 0.28”
Labconco P/N
3838500
3838501
3838502
3838503
Motor/Blower Data
Labconco P/N
6
3832200
3832201
3832207
3832208
Motor HP
½
½
¾
¾
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F25T8 TL741 F32T8 TL741
F40T8 SP41
F40T8 SP41
Fluor. Lamp Labconco Part #
9721901
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Ventus Canopy Data
Labconco Canopy P/N
3889200
3889201
3889202
3889203
Canopy Slot Area (m
2
)
.0111
.0111
.0111
.0111
Canopy Inflow Range (m/s)
.025-.047
.025-.047
.025-.047
.025-.047
Nominal Canopy Slot Velocity (m/s)
1.22-1.32
1.22-1.32
1.22-1.32
1.22-1.32
Nominal Canopy Slot Volume (m3/s)
.025
.025
.025
.025
Canopy Vacuum (Pa)
25-37
25-37
25-37
25-37
Nominal Total Exhaust Volume -
C Canopy (m3/s)
7
.15
.19
.23
.27
Traverse Total Exhaust Volume -
C Canopy (m3/s)
8
.17
.21
.25
.30