Chapter 5:
Using the Nano Enclosure
Now that the installation of your Nano enclosure is completed, you are ready to
use your Nano enclosure.
Read this chapter to learn about:
Routine Daily Work Procedures
Appropriate ULPA Filter Applications, Suitability and Guidelines
Optional Ionizer
Prohibited Acid Use
Routine Daily Work Procedures
Planning
Arrange for minimal disruptions, such as room traffic or entry into the
room while the enclosure is in use.
Start-up
Turn on exhaust system and light.
Only raise the sash for loading and cleaning.
Check the baffle air slots for obstructions.
Allow the enclosure to operate unobstructed for 1 minute.
Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.
Wear a protective mask if appropriate. Consult your Safety Officer for
additional personal protective equipment recommendations.
Loading Materials and Equipment
Load only the materials required for the procedure. Do not overload the
enclosure.
Do not obstruct the air foil, or rear baffle slots.
Large objects should not be placed close together and should be elevated
above the work surface to permit airflow to sweep under the equipment.
After loading, wait one minute to purge airborne contaminants from the
work area.