background image

Labnet Incubator Shaker

I5311-DS and I-5311-DS-230V

Instruction Manual

PO Box 841 Woodbridge, NJ 07095 USA
Phone: 732-417-0700

Fax: 732-417-1750

www.labnetlink.com

Copyright July 2004

Summary of Contents for I5311-DS

Page 1: ...Labnet Incubator Shaker I5311 DS and I 5311 DS 230V Instruction Manual PO Box 841 Woodbridge NJ 07095 USA Phone 732 417 0700 Fax 732 417 1750 www labnetlink com Copyright July 2004 ...

Page 2: ...dicates Potential Shock Hazard Indicates Protective Earthground SAFETY PRECAUTIONS Use of this product in any manner not specified by the manufacturer may impair the safety protection provided by the equipment and may result in physical damage and or personal injury Please read all operating instructions in this manual prior to use of this equipment 1 Do not operate this unit in an explosive or fl...

Page 3: ... may be purchased The I 5311 DS shaking mechanism supports a wide range of optional rack and flask platforms which are userchangeable SPECIFICATIONS Chamber temperature range Ambient 5C to 70C Ambient 10 C with shaker on Accuracy Uniformity 37 C 0 3C 0 5C Temperature accuracy 0 2C TemperatureController Microprocessor User Calibratable Over temperature safety Independentlysettable Shaker speed 20 t...

Page 4: ...aker shut off switch and a chamber light switch Main Power Switch This switch illuminated GREEN when on turns the power to the unit On and Off Temperature Controller and Temperature Set The controller has a 3 digit display for displaying chamber temperature or set point information UP and Down arrow pads are used to change the set point and controller mode of operation To enter the set point mode ...

Page 5: ...lamp illuminates when the shaker display is speed display mode showing the current shaking speed or showing the set shaking speed Stop Start This pad starts and stops the shaking mechanism and causes the RUN lamp to be illuminated Push to Shift Turn Set Knob This encoder provides the following multiple functions It increases and decreases the shaker speed setting in 1 rpm increments from 20 rpm to...

Page 6: ...et temperature before using Once the temperature shaking speed and shaking time are set use the Stop Start pad to initiate shaking When shaking in the timed mode the word End will be displayed on the shaker display when the shaking is done In timed mode the word Hld hold will be displayed on the shaker display The user may switch the shaker display to show the shaking speed by pressing the Push to...

Page 7: ...will not heat above temperature Confirm set point Check temperature of chamber that is below set point with a thermometer and recalibrate if needed Temperature display and reference Be sure that unit has been allowed to stabilize for thermometer do not match 1 hour Thermometer should be suspended in thechamberandnottouchinganysurface Only certified reference thermometers should be used Cannot adju...

Page 8: ...atform I 5330 500 8 x500ml flasks platform I 5330 1000 6 x 1000ml flasks platform I 5330 2000 4 x 2000ml flasks platform Flask clamps are available in 50ml 125ml 150ml 500ml 1000ml and 2000ml Test tube rack clamps are also available Labnet is continuously adding accessories Check with your representative or call Labnet directly at 732 417 0700 or check our website www labnetlink com for a complete...

Reviews: