INTRODUCTION
This manual covers the specifications, operation and use of the Labnet I-5311-DS shaking incuba-
tor and its accessories. Please pay special attention to the
Safety Precautions
section in this
manual.
The I-5311-DS shaking incubator from Labnet International provides an extremely stable tempera-
ture environment, and an aggressive shaking motion for the mixing of materials for cell culture
growth at ambient and above ambient temperatures.
The I-5311-DS shaking incubator uses microprocessor controls and mechanical convection to
maintain a stable temperature environment and to achieve fast chamber temperature recovery
after a door opening. The incubator interior is constructed from stainless steel for corrosion resis-
tance and easy clean-up of accidental spills. All incubator doors have an integral glass window and
the incubator has a light to allow observation of samples without opening the door. Each door is
fully thermal gasketed.
The I-5311-DS shaking incubator comes with one removable full shelf and four large, rubber pad-
ded adjustable feet for leveling the unit. Additional half and full shelves may be purchased. The I-
5311-DS shaking mechanism supports a wide range of optional rack and flask platforms which are
user changeable.
SPECIFICATIONS
Chamber temperature range
A5C to 70C (A10°C with shaker on)
Accuracy / Uniformity @37°C
+/-0.3C / +/-0.5C
Temperature accuracy
+/-0.2C
Temperature Controller
Microprocessor - User Calibratable
Over temperature safety
Independently settable
Shaker speed
20 to 300 rpm
Shaker orbit
19mm
Shaker max load
7.4kg*
Shaker timer
Continuous (Hld) or 30 seconds to 99 minutes 50 seconds
Timer increments
30 seconds to 9 min 59 seconds in 1 second steps
10.0 minutes to 99 min 50 seconds in 10 second increments
Interior electrical outlet
1 amp
Chamber volume
2.5 cu ft / 70 liters
Exterior dimensions w x d x h
574 x 544 x 644mm
Chamber dimensions w x d x h
442 x 396 x 406mm
Electrical Ratings and Unit Size
I-5311-DS
120V, 50/60Hz, 6 amps
I-5311-DS-230V
230V, 50/60Hz, 4 amps
* Max load is 4 x 2 liter flasks with 400ml liquid in each flask
UNPACKING
Upon receipt of your I-5311-DS shaking incubator, examine the carton and unit for damages. If shipping
damage has occurred, a claim must be filed with the carrier. The carrier is responsible for correcting
shipping damages. Save all packaging until the unit has been shown to operate properly to
your satisfaction. Carefully remove the unit from the carton and shipping pallet.