background image

INTRODUCTION

This manual covers the specifications, operation and use of the Labnet I-5311-DS shaking incuba-
tor and its accessories. Please pay special attention to the 

Safety Precautions

 section in this

manual.

The I-5311-DS shaking incubator from Labnet International provides an extremely stable tempera-
ture environment, and an aggressive shaking motion for the mixing of materials for cell culture
growth at ambient and above ambient temperatures.

The I-5311-DS shaking incubator uses microprocessor controls and mechanical convection to
maintain a stable temperature environment and to achieve fast chamber temperature recovery
after a door opening.  The incubator interior is constructed from stainless steel for corrosion resis-
tance and easy clean-up of accidental spills.  All incubator doors have an integral glass window and
the incubator has a light to allow observation of samples without opening the door. Each door is
fully thermal gasketed.

The I-5311-DS shaking incubator comes with one removable full shelf and four large, rubber pad-
ded adjustable feet for leveling the unit.  Additional half and full shelves may be purchased.  The I-
5311-DS shaking mechanism supports a wide range of optional rack and flask platforms which are
user changeable.

SPECIFICATIONS

Chamber temperature range

A5C to 70C (A10°C with shaker on)

Accuracy / Uniformity @37°C

+/-0.3C  /  +/-0.5C

Temperature accuracy

+/-0.2C

Temperature Controller

Microprocessor - User Calibratable

Over temperature safety

Independently settable

Shaker speed

20 to 300 rpm

Shaker orbit

19mm

Shaker max load

7.4kg*

Shaker timer

Continuous (Hld) or 30 seconds to 99 minutes 50 seconds

Timer increments

30 seconds to 9 min 59 seconds in 1 second steps
10.0 minutes to 99 min 50 seconds in 10 second increments

Interior electrical outlet

1 amp

Chamber volume

2.5 cu ft / 70 liters

Exterior dimensions w x d x h

574 x 544 x 644mm

Chamber dimensions w x d x h

442 x 396 x 406mm

Electrical Ratings and Unit Size

I-5311-DS

120V,  50/60Hz,  6 amps

I-5311-DS-230V

230V,  50/60Hz,  4 amps

* Max load is 4 x 2 liter flasks with 400ml liquid in each flask

UNPACKING

Upon receipt of your I-5311-DS shaking incubator, examine the carton and unit for damages.  If shipping
damage has occurred, a claim must be filed with the carrier.  The carrier is responsible for correcting
shipping damages.  Save all packaging until the unit has been shown to operate properly to
your satisfaction.  Carefully remove the unit from the carton and shipping pallet.

Summary of Contents for I5311-DS

Page 1: ...Labnet Incubator Shaker I5311 DS and I 5311 DS 230V Instruction Manual PO Box 841 Woodbridge NJ 07095 USA Phone 732 417 0700 Fax 732 417 1750 www labnetlink com Copyright July 2004 ...

Page 2: ...dicates Potential Shock Hazard Indicates Protective Earthground SAFETY PRECAUTIONS Use of this product in any manner not specified by the manufacturer may impair the safety protection provided by the equipment and may result in physical damage and or personal injury Please read all operating instructions in this manual prior to use of this equipment 1 Do not operate this unit in an explosive or fl...

Page 3: ... may be purchased The I 5311 DS shaking mechanism supports a wide range of optional rack and flask platforms which are userchangeable SPECIFICATIONS Chamber temperature range Ambient 5C to 70C Ambient 10 C with shaker on Accuracy Uniformity 37 C 0 3C 0 5C Temperature accuracy 0 2C TemperatureController Microprocessor User Calibratable Over temperature safety Independentlysettable Shaker speed 20 t...

Page 4: ...aker shut off switch and a chamber light switch Main Power Switch This switch illuminated GREEN when on turns the power to the unit On and Off Temperature Controller and Temperature Set The controller has a 3 digit display for displaying chamber temperature or set point information UP and Down arrow pads are used to change the set point and controller mode of operation To enter the set point mode ...

Page 5: ...lamp illuminates when the shaker display is speed display mode showing the current shaking speed or showing the set shaking speed Stop Start This pad starts and stops the shaking mechanism and causes the RUN lamp to be illuminated Push to Shift Turn Set Knob This encoder provides the following multiple functions It increases and decreases the shaker speed setting in 1 rpm increments from 20 rpm to...

Page 6: ...et temperature before using Once the temperature shaking speed and shaking time are set use the Stop Start pad to initiate shaking When shaking in the timed mode the word End will be displayed on the shaker display when the shaking is done In timed mode the word Hld hold will be displayed on the shaker display The user may switch the shaker display to show the shaking speed by pressing the Push to...

Page 7: ...will not heat above temperature Confirm set point Check temperature of chamber that is below set point with a thermometer and recalibrate if needed Temperature display and reference Be sure that unit has been allowed to stabilize for thermometer do not match 1 hour Thermometer should be suspended in thechamberandnottouchinganysurface Only certified reference thermometers should be used Cannot adju...

Page 8: ...atform I 5330 500 8 x500ml flasks platform I 5330 1000 6 x 1000ml flasks platform I 5330 2000 4 x 2000ml flasks platform Flask clamps are available in 50ml 125ml 150ml 500ml 1000ml and 2000ml Test tube rack clamps are also available Labnet is continuously adding accessories Check with your representative or call Labnet directly at 732 417 0700 or check our website www labnetlink com for a complete...