background image

-Heating Indicator

-This indicator will illuminate GREEN when the controller is calling for heat

from the heater.  This indicator will be on continuously while the oven heats up to the set tem-
perature and will then cycle on and off at the set temperature.

-Safety Thermostat

-The safety thermostat located on the back of the unit is manually set and

completely independant of the Main Termperature Controller.  The safety gurads against any
failure of the Main Controller that would allow temperature to rise past the safety set point.  If
the temperature rises to the safety set point, the Safety takes control of the heating element
and allows continued use of the incubator until the problem can be resolved or service can be
arranged.

-Safety Indicator

-This indicator wil illimunate RED when the Safety Thermostat is activated.

Under normal operating conditions this indicator should never be on.

-Chamber Light Switch-

This switch turns the chamber light on and off when power is on.

-SHAKER CONTROLS-

The shaker controls are also microprocessor based.

-RUN lamp-

This lamp illuminates when the shaker is started and stays illuminated when the

shaker is running.

-Timer lamp-

This lamp is illuminated when the shaker display is in the time display mode

counting down the remaining run time or showing the set run time.

-RPM lamp-

This lamp illuminates when the shaker display is speed display mode showing the

current shaking speed or showing the set shaking speed.

-Stop / Start-

This pad starts and stops the shaking mechanism and causes the RUN lamp to

be illuminated.

-Push to Shift / Turn Set Knob- 

This encoder provides the following multiple functions:

It increases and decreases the shaker speed setting in 1 rpm increments from 20 rpm
to 300 rpm.

It increases and decreases the run time setting from 30 seconds to 99 minutes and 50
seconds and HOLD (Hld) for continuous operation.  Time is changed in 1 second
increments to 10 minutes and then 10 second increments thereafter.

It shifts the shaker display from time display to speed display and back

-Shaker Display- 

This three digit display shows the shaker speed setting or shaker time

setting or remaining shaker run time while counting down.  It also shows:

Hld 

- When the shaker is set for continuous un-timed running

Opn

 - When the incubator door has been opened, automatically stopping the shaking

motion (note the timer continues to count down but is not displayed)

End

 - When a timed shaking run has ended

F60 / F50

 - When the shaking incubator is first powered up indicating the line

frequency of the power line.

-Door Open Switch - 

This switch senses when the incubator shaker door is open and sends

a signal to the microprocessor which causes the shaker to stop running as a safety measure.
However, in a timed run, the timer will continue to count down when the door is open.  Closing
the door restarts the shaking motion.

Summary of Contents for I5311-DS

Page 1: ...Labnet Incubator Shaker I5311 DS and I 5311 DS 230V Instruction Manual PO Box 841 Woodbridge NJ 07095 USA Phone 732 417 0700 Fax 732 417 1750 www labnetlink com Copyright July 2004 ...

Page 2: ...dicates Potential Shock Hazard Indicates Protective Earthground SAFETY PRECAUTIONS Use of this product in any manner not specified by the manufacturer may impair the safety protection provided by the equipment and may result in physical damage and or personal injury Please read all operating instructions in this manual prior to use of this equipment 1 Do not operate this unit in an explosive or fl...

Page 3: ... may be purchased The I 5311 DS shaking mechanism supports a wide range of optional rack and flask platforms which are userchangeable SPECIFICATIONS Chamber temperature range Ambient 5C to 70C Ambient 10 C with shaker on Accuracy Uniformity 37 C 0 3C 0 5C Temperature accuracy 0 2C TemperatureController Microprocessor User Calibratable Over temperature safety Independentlysettable Shaker speed 20 t...

Page 4: ...aker shut off switch and a chamber light switch Main Power Switch This switch illuminated GREEN when on turns the power to the unit On and Off Temperature Controller and Temperature Set The controller has a 3 digit display for displaying chamber temperature or set point information UP and Down arrow pads are used to change the set point and controller mode of operation To enter the set point mode ...

Page 5: ...lamp illuminates when the shaker display is speed display mode showing the current shaking speed or showing the set shaking speed Stop Start This pad starts and stops the shaking mechanism and causes the RUN lamp to be illuminated Push to Shift Turn Set Knob This encoder provides the following multiple functions It increases and decreases the shaker speed setting in 1 rpm increments from 20 rpm to...

Page 6: ...et temperature before using Once the temperature shaking speed and shaking time are set use the Stop Start pad to initiate shaking When shaking in the timed mode the word End will be displayed on the shaker display when the shaking is done In timed mode the word Hld hold will be displayed on the shaker display The user may switch the shaker display to show the shaking speed by pressing the Push to...

Page 7: ...will not heat above temperature Confirm set point Check temperature of chamber that is below set point with a thermometer and recalibrate if needed Temperature display and reference Be sure that unit has been allowed to stabilize for thermometer do not match 1 hour Thermometer should be suspended in thechamberandnottouchinganysurface Only certified reference thermometers should be used Cannot adju...

Page 8: ...atform I 5330 500 8 x500ml flasks platform I 5330 1000 6 x 1000ml flasks platform I 5330 2000 4 x 2000ml flasks platform Flask clamps are available in 50ml 125ml 150ml 500ml 1000ml and 2000ml Test tube rack clamps are also available Labnet is continuously adding accessories Check with your representative or call Labnet directly at 732 417 0700 or check our website www labnetlink com for a complete...