User should take care of two major points, if especially the recrystallized tube size is 
larger than 60mms.

Programming should be applied to manage controlled heating to set temperature 

and also controlled cooling is strongly recommended. Heat up rates of less than 
8°C/min is strongly advised.

Always avoid removal of any substance within the tube while hot as this will create 

a sudden cool down at point of contact. If absolutely necessary, use a insulating 
sample carrier plate where the sample is placed upon during heating stage.

Intervention into the tube ,placing a sample while hot can also be a reason for 

cracking, always use a carrier base plate

If relatively large samples are placed within the tube always take care for a slower 

heat up to enable a stable heating.

5.

We recommend the use of furnaces with SiC heating elements without any shut downs, 

kept at stand by at around 4000C when unoperational. This procedure will increase the 
service life of SiC heating elements. Furnaces with wire heating do not require any such 
precautions and can be turned off whenever required.

Whenever a breakage occurs of SiC elements, user has to change it in pairs, keeping the 

unbroken one in stock. It is not possible to use an old SiC element together with a new 
one.

The old unbroken element should be kept aside to be used in pairs with an old element.

At  high  temperatures  a  glassy  phase  coating  appears  on  SiC  elements. This  coating 

although not harmful to elements may fracture and disintegrate to small pieces of glass if 
handled.  For  reasons  of  safety  use  a  glove  and  eye  glass  while  handling  old  SiC 
elements. 

6.

For PTF series furnaces of over 1400°C, a standard over temperature alarm system is 

employed to protect the SiC elements. If ever over temperature increase occurs, this alarm 
system cuts off the power by means of disengaging a contactor. This function is activated by 
the small knob placed on the panel which has to be pressed at initial start ups.

So long as the power is not turned down repeated pressing is unnecessary as O/T 
alarm system will always be active. Furnaces with this system will only start heating 
when this knob is pressed. You will hear the sound of a contactor closing when the 
knob is pressed.

7.

Keep organic solvents and water away from insulation material. 

8.

Keep metal slag away from heating elements to avoid a short cut.

9.

Labnics furnaces are not designed for heating food and operator should not use the furnace 

for cooking purposes.

10.

Heating elements usually give a long service life but are basically for consumption. They will 

have to be replaced after a reasonable period of use. Heating elements are not a part of 
Warranty which is provided for a period of 2 years.

i

i

Summary of Contents for LSCF-100A

Page 1: ...Instruction Manual Model LHTF 100 Series Labnics Equipment High Temperature Tube Furnace Please read this manual carefully before using the instrument...

Page 2: ...Technical Properties 4 2 1 Temperature Controllers 4 2 2 Heating Elements 7 2 3 Refractory Insulation 8 2 4 Electrical Layout 8 3 What is PID Control 9 4 Points to Take Care While Operating the Furnac...

Page 3: ...serts slowly It is extremely important to remove the packing inserts from inside the furnace Operating the furnace without removal of safety inserts will cause heavy damage on the heating elements Use...

Page 4: ...b requires a qualified technician The technician who is responsible of commissioning the furnace is also responsible to act in accordance the rules of safety of European Community Technician is solely...

Page 5: ...d against risks to avoid any burns etc 6 Labnics furnaces are manufactured in conformity with CE Electromagnetic Emissions Directive Even though so electromagnetic emissions can effect some sensitive...

Page 6: ...tured at will and without informing 2 Furnace Components Related Technical Properties 2 1 Temperature Controllers Model Sensitivity PID Step definition Number of Number of Programs Segments DC 1010 1...

Page 7: ...active functions of the controller Insignia Description Active Color OUT1 OUT2 Output Relays power are active Green AT AutoTune is underway Orange AL1 AL2 AL3 Alarm signal underway Red MAN Controller...

Page 8: ...play any programming profile has to have the cooling stage defined as well Use a paper to draw your program profile before data feed in We advice the PVSTART option to be enabled to ease the running o...

Page 9: ...y Used to either to decrease a value or to inactivate a function 2 2 Heating Elements Labnics employs Kanthal originated heating elements in furnaces 2 2 1 Wire elements The basic advantages of Kantha...

Page 10: ...s of service as a heating element The resistance of Kanthal Super elements does not change due to ageing even after having been in operation for a long time at high temperatures Due to these propertie...

Page 11: ...erature and it should never be used over this temperature The heating elements and the concerned insulation are related with designated o maximum temperature of furnace It is most advicable to us a fu...

Page 12: ...a new one The old unbroken element should be kept aside to be used in pairs with an old element At high temperatures a glassy phase coating appears on SiC elements This coating although not harmful to...

Page 13: ...use insulation that appears loose Most of the time the very same insulation can be obtained from local market We can of course deliver the concerned insulation whenever requested B Replacement of Sol...

Page 14: ...ently asked questions FaultAnalysis 1 Furnace controller does not display when the I 0 switch turned on 1 1 Check the Mains supply Check the voltage present May be less than 220V 1 2 Check the fuses o...

Page 15: ...1 Electrical Other Requirements Phase Volts Frequency Medium Temp Humidity Single 220V 50Hz 5 C 40 C Max 85 Triphase 380 220V 50Hz 2 Control Panel Honeywell DC1010 UDC2500 Honeywell DC1010 Honeywell U...

Page 16: ...igits can be activated by pressing key By using and keys you can define the digits of the desired limit temperature After you have entered the limit temperature value please press set key in order to...

Page 17: ...ys in the manual mode and open red the I 0 switch Furnace will start heating directly to set temperature 2 Programme Mode See Section C page 11 and Section D page 15 for further details Press key cont...

Page 18: ...well time has expired 2 Use or key to select the requested recepie program number Please note that program number changes as the operator press or keys 3 Once a programme is chosen in order to run the...

Page 19: ...rogramme data during operation E Sample Program Honeywell UDC2500 There are four different alternatives in order to heat up the furnace with UDC2500Temperature Controller 1 Standard Heating 2 To defin...

Page 20: ...o next step ENDSEG End Segment max 12 The segment number which the programe will end RPUNIT Engineering Unit for Ramp TIME Defines heating up or cooling down speed Can be chosen as TIME Hr Min EU H C...

Page 21: ...t point Unit is Hr Min SG3RP RampTime 3 1hr SG4SP Soak SP 4 400 SG4TI SoakTime 4 1hr SG5RP RampTime 5 1hr 30min SG6SP Soak SP 6 250 SG6TI SoakTime 6 3hr 0min SG7RP RampTime 7 2hr 30min SG8SP Soak SP 8...

Page 22: ...rts Replaced Parts Recommended Action Required Yes Service Engineer s Name Signature Requisition Number Customer s Name Signature Date Stamp Page ____ Of ____ No Model Serial No Date Status OK Install...

Page 23: ...Labnics Equipment 43040 Christy St Fremont CA 94538 USA Toll Free 877 620 9992 Tel 925 271 4322 Fax 925 886 0400 Email info labnics com Website www labnics com...

Reviews: