background image

A.

INTRODUCTION

This booklet includes our recommendations, hints and points of notice for you to have a safe and better service 
from this equipment.  We kindly ask you to carefully read this manual and only then to operate the furnace.

This insignia is used for the purpose of
warning you against risks involved. 
Read carefully the notes when this sing is noted.

The insignia is to warn you for any risk of 
damage to this equipment.

This insignia warns you of the hints to use 
this equipment in better terms.

B.

Information Related To Safety

1.

Unpacking and Handling of the Furnace 
Before operating the furnace please take note of following points.

This  furnace  has  been  packed  to  minimize  any  risk  due  to  transportation.  Remove  the 

packing with care; avoid using any kind of force and removing packing inserts slowly.

It is extremely important to remove the packing inserts from 
inside the furnace .Operating the furnace without removal 
of safety inserts will cause heavy damage on the heating elements.

Use furnace base while lifting and carrying the furnace.

Always use two persons to carry.

Always take note of the weight of furnace during placement.

Do not operate the furnace if there is inflammable material around. Ensure that the risk of 

fire is non-existent .Do not place the furnace on an inflammable surface.

Placement should be in a well ventilated room, with no interaction with other sources of 

heat. Do not place on an inflammable surface.

Air circulation, should be capable enough to remove any gas emission 
due to come through the lid or furnace chimney. Almost all gas 
emissions during operation can be hazardous for Human health 

Always give space of 30cm around the furnace to enable quick service. Do not close the 
vents on the furnace body. These are required to cool down the system.

- 1 -

i

i

i

Summary of Contents for LSCF-100A

Page 1: ...Instruction Manual Model LHTF 100 Series Labnics Equipment High Temperature Tube Furnace Please read this manual carefully before using the instrument...

Page 2: ...Technical Properties 4 2 1 Temperature Controllers 4 2 2 Heating Elements 7 2 3 Refractory Insulation 8 2 4 Electrical Layout 8 3 What is PID Control 9 4 Points to Take Care While Operating the Furnac...

Page 3: ...serts slowly It is extremely important to remove the packing inserts from inside the furnace Operating the furnace without removal of safety inserts will cause heavy damage on the heating elements Use...

Page 4: ...b requires a qualified technician The technician who is responsible of commissioning the furnace is also responsible to act in accordance the rules of safety of European Community Technician is solely...

Page 5: ...d against risks to avoid any burns etc 6 Labnics furnaces are manufactured in conformity with CE Electromagnetic Emissions Directive Even though so electromagnetic emissions can effect some sensitive...

Page 6: ...tured at will and without informing 2 Furnace Components Related Technical Properties 2 1 Temperature Controllers Model Sensitivity PID Step definition Number of Number of Programs Segments DC 1010 1...

Page 7: ...active functions of the controller Insignia Description Active Color OUT1 OUT2 Output Relays power are active Green AT AutoTune is underway Orange AL1 AL2 AL3 Alarm signal underway Red MAN Controller...

Page 8: ...play any programming profile has to have the cooling stage defined as well Use a paper to draw your program profile before data feed in We advice the PVSTART option to be enabled to ease the running o...

Page 9: ...y Used to either to decrease a value or to inactivate a function 2 2 Heating Elements Labnics employs Kanthal originated heating elements in furnaces 2 2 1 Wire elements The basic advantages of Kantha...

Page 10: ...s of service as a heating element The resistance of Kanthal Super elements does not change due to ageing even after having been in operation for a long time at high temperatures Due to these propertie...

Page 11: ...erature and it should never be used over this temperature The heating elements and the concerned insulation are related with designated o maximum temperature of furnace It is most advicable to us a fu...

Page 12: ...a new one The old unbroken element should be kept aside to be used in pairs with an old element At high temperatures a glassy phase coating appears on SiC elements This coating although not harmful to...

Page 13: ...use insulation that appears loose Most of the time the very same insulation can be obtained from local market We can of course deliver the concerned insulation whenever requested B Replacement of Sol...

Page 14: ...ently asked questions FaultAnalysis 1 Furnace controller does not display when the I 0 switch turned on 1 1 Check the Mains supply Check the voltage present May be less than 220V 1 2 Check the fuses o...

Page 15: ...1 Electrical Other Requirements Phase Volts Frequency Medium Temp Humidity Single 220V 50Hz 5 C 40 C Max 85 Triphase 380 220V 50Hz 2 Control Panel Honeywell DC1010 UDC2500 Honeywell DC1010 Honeywell U...

Page 16: ...igits can be activated by pressing key By using and keys you can define the digits of the desired limit temperature After you have entered the limit temperature value please press set key in order to...

Page 17: ...ys in the manual mode and open red the I 0 switch Furnace will start heating directly to set temperature 2 Programme Mode See Section C page 11 and Section D page 15 for further details Press key cont...

Page 18: ...well time has expired 2 Use or key to select the requested recepie program number Please note that program number changes as the operator press or keys 3 Once a programme is chosen in order to run the...

Page 19: ...rogramme data during operation E Sample Program Honeywell UDC2500 There are four different alternatives in order to heat up the furnace with UDC2500Temperature Controller 1 Standard Heating 2 To defin...

Page 20: ...o next step ENDSEG End Segment max 12 The segment number which the programe will end RPUNIT Engineering Unit for Ramp TIME Defines heating up or cooling down speed Can be chosen as TIME Hr Min EU H C...

Page 21: ...t point Unit is Hr Min SG3RP RampTime 3 1hr SG4SP Soak SP 4 400 SG4TI SoakTime 4 1hr SG5RP RampTime 5 1hr 30min SG6SP Soak SP 6 250 SG6TI SoakTime 6 3hr 0min SG7RP RampTime 7 2hr 30min SG8SP Soak SP 8...

Page 22: ...rts Replaced Parts Recommended Action Required Yes Service Engineer s Name Signature Requisition Number Customer s Name Signature Date Stamp Page ____ Of ____ No Model Serial No Date Status OK Install...

Page 23: ...Labnics Equipment 43040 Christy St Fremont CA 94538 USA Toll Free 877 620 9992 Tel 925 271 4322 Fax 925 886 0400 Email info labnics com Website www labnics com...