6.2.3
Set-point pressure
Touch screen
– as opened
Item
Action
÷
Decrease set-point by 0,001 hPa/Torr.
Accept change in set-point.
+
Increase set-point by 0,001 hPa/Torr.
Press the LaboGene logo to return to the
manual menu.
Press the LaboGene log for 2-3 seconds
to lock the display for input. Unlock press
again.
6.2.4
Alarm pressure
Touch screen
– as opened
Item
Action
The alarm pressure is a limiting parameter that ensures that the
shelf heating is not turned on before this pressure is reached.
If the pressure raise above this pressure, during operation, the
shelf heating will be turned off to protect the product being
freeze dryed.
÷
Decrease set-point by 0,1 hPa/Torr.
Accept change in set-point.
+
Increase set-point by 0,1 hPa/Torr.
Press the LaboGene logo to return to the
manual menu.
Press the LaboGene log for 2-3 seconds
to lock the display for input. Unlock press
again.
6.3
Auto menu
Touch screen
– as opened
Item
Action
Active programme, press to select
programme from list.
Start active programme.
Edit active programme.
Press the LaboGene logo to return to the
main menu.
Press the LaboGene log for 2-3 seconds
to lock the display for input. Unlock press
again.