Code 9
9.900.900.763
5/19
U:\QA\Manualer\Arkiv\Manuals- Labogene-Origio-GKO\Manualer-WORD\ScanVac\CoolSafe
XS
XL Usermanual-Rev A5.docx
It is advisable that Safety glasses and personal protective garments be worn
by all User’s of the equipment. CoolSafe XS & XL Models are fitted with
castors, the front pair being lockable. Once located in the position conducive
to both the safety of the laboratory environment and personnel, it is advised
that the castors be in the locked position.
The safety system is designed according to Safety requirements for electric
equipment, control and laboratory use (EN 61010-1), which assumes that
two separate faults do not occur simultaneously.
3. Installation
3.1.
Installation of the CoolSafe
Make sure that the mains supply is according to the specifications on the
type label at the rear panel.
It is important that free air circulation is available at the air inlet and outlet
grills. If the CoolSafe is positioned against a wall, there must be at least 100
mm free space at the rear of the unit.
CoolSafe 100-9 PRO Superior XS & CoolSafe 95-15 PRO Superior XL are both
floor standing units fitted with castors, the front pair being lockable. Before
operating the system insure that the castors are in the locked position.
3.2. Installation of vacuum pump
A two stage vacuum pump (with 2-5 m
3
/hour capacity or greater) must be connected
to the CoolSafe via the connection on the rear of the CoolSafe in order to allow cryo
pumping.
The vacuum pump can be located on the rear panel of the CoolSafe by means of a
shelf support. This is located by using the thumb screws provided-see below.