Prima Mµ
6210000-795
Prima Mµ
Issue 1.6
Printed on Jan.,2021
10
8. ELECTRICAL CONNECTIONS
Fig. 7
Fig. 8
Fig. 9
Device Name
Model Number
Serial Number
Input Voltage
Frequency
Max. Load
Separate Disposal
waste electric device
Manufacturer
Company
Trademark
Label Control
Number
Fuse Rating
IEC60601-1:2005
Labeling for Canada
IEC60601-1:2005
Labeling for USA & Europe
Accompanying Document
must be consulted
2
1
3
A
AC Inlet
On/Off Switch
Mu
W
Connect the power cable to the AC inlet socket
(2) provided on the back of the swivel arm as
shown in the fig.-7.
Switch on the power from on/off switch (1).
Note: Power supply is designed with universal
input 100V - 240V AC, 50/60HZ. To plug in follow
Instruction on electrical label provided at bottom
of the arm as shown (3) here in Fig.-7.
8.1 REPLACING THE ILLUMINATION SOURCE
Open the swivel arm cover’s. Detach the fibre
optic cable and replace the illumination assembly
A with new assembly. Secure back the arm cover’s.
8.2 FUSE REPLACEMENT
The fuse is located with the AC inlet, provided near
the on/off switch. Use a flat head screw driver to
open the fuse compartment. Two fuses are provided
in this, i.e. one is live fuse and second as spare fuse.
Replace the blown fuse with live fuse and secure
back the fuse compartment. For fuse replacement refer
label shown in fig.-9.
8.3 DATA PLATE FOR PRIMA Mµ MICRO-
SCOPE