Prima Mµ
6210000-795
Prima Mµ
Issue 1.6
Printed on Jan.,2021
28
GUIDANCE TABLES (continued)
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Max Output Power of
Transmitter (W)
Separation (m)
150kHz to 80 MHz
d=(3.5/v1) (Sqrt P)
Separation (m) 80 to
800 MHz
d= (3.5/E1) (Sqrt P)
Separation (m) 800
MHz to 2.5GHz
d= (7/E1) (Sqrt P)
0.01
0.1
1
10
100
0.1166
0.3689
1.1666
1.1666
3.6893
3.6893
11.6666
11.6666
0.1166
0.3689
0.2333
0.7378
2.3333
7.3786
23.3333
Recommended separation distance between Portable and Mobile RF Communications equipment
and the Prima Mµ for ME equipment and ME systems that are not Life-supporting.
Recommended separation distance for between Portable and Mobile RF Communications equipment
and the Prima Mµ
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: At 80MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
The Prima Mµ is intended for use in electromagnetic environments in which radiated RF disturbances
are controlled. The customer or user of the Prima Mµ can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Prima Mµ as recommended below, according to the maximum output power of the
Communication equipment.