II) BULB REPLACEMENT
III) INSTALLING OR REPLACING THE FUSE
1. Before attaching the lamp bulb, remove the parts that may drop
such as the filter and specimen from the Microscope frame, and
place the Microscope on its back so that the bottom plate is
exposed.
2. Pull the lock knob (1) on the bottom to open lamp housing door
(fig.1).
3. Hold the Halogen bulb (2) without taking it out of the polyeth-
ylene bag so as not to taint the bulb with fingerprints and push
the bulb into the pin holes on the socket (3). After attaching,
remove the polyethylene bag.
4. With the lock knob pilled out, close the lamp housing door, then
push the lock knob back to lock the cover. Always use the desig-
nated bulb. Using a bulb other than those specified by LABOMED
may lead to a fire hazards. Fingerprints or stains on the lamp
bulb reduce its life. If contamination occurs, wipe bulb surface
with a cloth slightly moistened with alcohol.
Applicable Bulb: 6V20W Halogen Bulb P/N CX-013
Caution: For Fuse Replacement
Set the main switch to “O” (OFF), disconnect the power cord from
the wall outlet.
Before replacing the fuse, remove the parts that may drop such as
the filter and specimen from the microscope frame. Turn around
the microscope to its back so that the AC inlet is visible.
1. Use a flat head screw driver to open the fuse holder (1).
2. The fuse tray will come out with (2) live fuse and (3)spare fuse.
Do not pull out the fuse tray with force as it is locked and will
not be out completely.
3. Replace the primary fuse (2) with the spare fuse.
4. Engage the fuse tray back in.
Always use the designated Fuse. Using a fuse other than those
specified by LABOMED may lead to a fire hazards.