Under certain conditions, performance of the unit may be adversely affected by factors other than defects.
If problems occur, please review the following list and take remedial action as needed. If you cannot solve
the problem after checking the entire list, please contact Labomed for assistance.
OBSERVATION CAUSE REMEDY
Uneven brightness in
observation field
All LED’s are not glowing in
incident light.
The LED does not glow.
The specimen does not focus
The incident illumination is low.
The transmitted illumination is
blocked by the specimen.
The LED control is wrongly select.
The power cord id not attached.
The LED are not switched on.
The power button is not switched
on.
The fuse is blown.
The specimen is not focused
properly.
The specimen is too small.
The specimen is too big.
Increase the incident illumination
Place an opaque specimen or move
the specimen. Use incident light.
Press the button to glow all LED’s,
11 front LED’s or 12 rear LED’s.
Check the power cord. Attach is
Precisely.
Switch on the LED’s.
Switch indicator will glow
when the power is on.
Check the fuse. If blown, replace
with a live fuse. Adhere caution!!
Focus the specimen first on the
lowest magnification. Achieve best
focus using the coarse adjustment
knob.
The specimen is too small. Slide
with micro dissection are not
visible on stereo microscopes. The
Specimen is too big to focus. Use
smaller specimen.
TROUBLESHOOTING GUIDE