126
Nexam Pro User Manual
Chapter 9
anal canal at a constant speed. A metronome function can be activated in the
protocol to aid in withdrawing the catheter. You have to set the pull
(withdrawal) speed and length (distance). During the investigation, each step
will be displayed on the screen at the set speed. The metronome makes a
sound at each step. The software automatically places the profile end marker.
9.3
Prepare the Investigation
9.3.1
Water Catheters
For measuring with water catheters and pressure transducers, perfusion of the
channels is needed (in case a perfusion pump is available (HRAM option)
please refer to the HRAM U
ser’s Manual for the preparation of the
investigation, LBL-001028).
The following items are included for perfusion with a pressure-cuff:
•
Pressure-cuff (300 mmHg).
•
DT-NN pressure transducers with built-in flow-reducers (4x).
•
CIM module.
•
4-way catheter flushing set.
Figure 9.15
Perfusion with pressure-cuff