128
Nexam Pro User Manual
Chapter 9
The transducers must be left in the OPEN position.
The procedure to prepare the investigation is as follows:
▪
Connect the pressure transducer cables (in OPEN position) to the CIM
module. For the GIM-60000A air-charged catheter and transducers, the
connection is as follows:
Color pressure transducer
CIM module
Yellow (distal)
P1
Green (medial distal)
P2
Blue (medial proximal)
P3
Orange (proximal)
P4
▪
Enter and select the
patient name
in the database program.
▪
Click the
New investigation
button in the database program.
▪
Click the
Anorectal manometry
button to start the pre-test.
▪
Connect the catheter to the pressure transducers (follow the colors). Be
sure to connect in the OPEN position.
▪
Calibrate the catheter as instructed by the manufacturer (follow the
instructions to calibrate the catheter and charge the sensors). Press the
Zero
all
button before charging the sensors in the calibration tube.
▪
Release pressure from the calibrator and remove the catheter. After
calibration, do not slide the switch to the OPEN position.
▪
Insert the catheter gently in the anal canal of the patient and position the
balloon in the rectum.
Figure 9.17
Charge catheter
▪
Ask the patient to cough to check the registration of the pressures.