Anorectal Manometry
Nexam Pro User Manual
137
Insert more of the same sensation markers
It is possible to perform the sensation test more than once and thus insert
more than one of the same type of sensation marker.
To reset (restart) the sensation test, press the
Next test
button until the
text
No sensation
is displayed on the left button, or
Right-click this button until the sensation marker you want to insert is
displayed.
9.4.9
Rest and Squeeze Profiles Using a Puller
The purpose of the anal resting and squeeze pressure profiles is to obtain a
continuous pressure profile of the entire anal sphincter. This is accomplished
by withdrawing the catheter through the anal-sphincter by means of the puller
or manually. In this section, you will find a description of the test using the
puller.
When pressures are measured radially throughout the sphincter, the software
can create cross-sectional and longitudinal graphs of the sphincter (3D vector
volume plots). The anal channels should all be marked with sensor positioning
Anal
in the catheter definition (see § 9.5.2).
Prepare the catheter puller and position the catheter on the puller as described
in § 2.5.12. The pressure channels of the catheter should be positioned in the
rectum.
NOTE
Do not position the puller too close to the anal verge. This
prevents the puller head from pollution by the catheter
during withdrawing.
The procedure is as follows:
Press the
Next test
button until the text
Start puller [Rest]
is
displayed on the left button.
Press the
Start puller [Rest]
button to start the puller. After a short
while there should be an increase in the anal pressure.
Summary of Contents for Nexam Pro CIM
Page 1: ...Nexam Pro User Manual...