34
Nexam Pro User Manual
Chapter 2
Figure 2.13
Transparent view of pump unit
1. Lever to press the tube clamp over the tube
2. Mark on tube clamp to indicate 5-roller or
10-roller pump
3. Tube clamp
4.
Rollers
5.
Control panel
6.
Flush button
7.
Weighing hook (infused volume transducer)
Install the pump tube as follows:
▪
Start the pre-test of the investigation.
▪
Turn the lever of the pump [1] and take out the tube clamp [3].
▪
Unpack the pump tube.
▪
Put the spike of the pump tube into the bag of saline.
▪
Hang the bag of saline on the hook [7] of the pump unit.
▪
Put the pump tube on the rollers; the tube has a special part which must be
put on the rollers, indicated by two stoppers.
Figure 2.14
Nexam Pro pump tube with two stoppers
Stopper
Stopper
Summary of Contents for Nexam Pro CIM