206
Nexam Pro User Manual
Chapter 12
12.5.9 Connections Nexam Pro
For each investigation, you can set the channel numbers for the channels
(signal) to be measured with the Nexam Pro measurement system. In the
protocol settings
window, select the investigation tab and then click the
Connections
button to set the connections for that investigation. The window
shown in Figure 12.16 opens.
Figure 12.16
Connections Nexam Pro, Channel configuration
In the left column, you will see the channels that have been selected via the
Channels
button (see § 12.5.8). For each channel (signal) to be measured,
you must select the channel number.
In this example, P
ves
is measured on channel 1 of the CIM module.
P
abd
is measured on channel 2. P
ura
is measured on channel 3.
EMG is measured on the EMG channel (1) on the CIM module.
It is possible to use more than one Bluetooth flowmeter. By default, the Nexam
Pro measurement system will use all connected or registered flowmeter(s) with
the
automatic selection
setting. You can also select one specific flowmeter.
When a flowmeter or puller is not used for a specific measurement (i.e.,
cystometry), you can use the option
not measured
for that investigation. In
this case, the software will not try to connect to this Bluetooth device.