Nexam Pro User Manual
Chapter 6
the buzzer of the remote control will beep after the flow has stopped.
(You can find more information in § 12.3.4).
Press the
Stop
investigation
button. Enter the residual urine if
prompted.
Remote control buttons
Buttons
Description
Press
Shift
and
Zero
all
simultaneously to zero the
detrusor pressure (P
det
) during the pre-test and, if
necessary, during the measurement.
Press
Valsalva begin/end
to start/end the Valsalva
maneuver.
Click
Spontaneous Miction
to mark spontaneous miction.
Press
Shift
and
Leakpoint
to start the Cough Leak Point
Pressure procedure (CLPP). Ask the patient to cough; the
software will insert cough markers around the peaks.
Press
Leakpoint
to mark a leakpoint.
6.4
Results
In the analysis program, choose
Results
>
Display results
to display the
calculated parameters. Figure 6.1 shows the parameters that can be
calculated.