- 25 -
8.1.8.
Animal and Dynamic Weighing
1.
To configure the Animal and Dynamic Weighing Function, use the navigation keys to select
“Animal
Weighing”
in
“Mode Setup”
and press ENTER. The status and available Animal Weighing options
will be displayed. If the function is currently in use the status will be set to
“ON”.
2.
To select an operating mode use the navigation keys to select
“Mode”
and press ENTER. By default
Animal Weighing is set to
“Auto”
(Automatic). To operate Animal Weighing in Semi-Automatic or
Manual use the navigation keys to select a desired operating mode and press ENTER.
Animal Weighing Operating Modes
Automatic
Weighing Initialization : Automatic
Result Clearing : Automatic
Tare : Automatic
When set to Automatic, the scale will
automatically begin the animal weighing
process. Once the animal is removed from
the weighing pan, the scale will
automatically tare and prepare for the next
weighing.
Semi- Automatic
Weighing Initialization : Automatic
Result Clearing : Manual
Tare : Manual
When set to Automatic, the scale will
automatically begin the animal weighing
process. Once the animal is removed from
the weighing pan, the result remains
displayed until it is manually cleared. The
scale must be manually tared before the next
weighing.
Manual
Weighing Initialization : Manual
Result Clearing : Manual
Tare : Manual
When set to Manual, the animal weighing
process must be manually initialized. Once
the animal has been weighed and the pan
cleared the result must be manually cleared.
The scale must be manually tared before the
next weighing.