S
YSTEM
M
AINTENANCE
INITIAL CARTRIDGE INSTALLATION
Carefully read and understand the full written instructions prior to installing cartridges.
This unit was designed to work in optimal condition with genuine LabStrong™ cartridges.
STEP 1|
Press the MENU button.
STEP 2|
Select and press the MAINTENANCE BUTTON.
STEP 3|
Select and press the INSTALL NEW CARTRIDGE button.
STEP 4|
Press the BEGIN button.
STEP 5|
Select and press the proper cartridge kit being installed.
STEP 6|
View the install order then press the NEXT button.
STEP 7|
Install the proper cartridge in the first position, (start right to left). Press NEXT.
STEP 8|
Make sure the tubes and fittings are installed properly then press the NEXT button.
STEP 9|
Install the remaining three cartridges in the proper order. Press the NEXT button.
STEP 10|
Make sure the bottom tubes and fittings are properly installed. Press the NEXT button.
STEP 11|
Once the cartridges are properly installed and the drain tube is installed at the dispenser port, turn
on the inlet water valve then press the PURGE AIR button.
STEP 12|
Once the air purge is completed, install the 0.2 Micron Final Filter and press the NEXT button.
STEP 13|
Replace the cover then press the NEXT button.
STEP 14|
Once installation is complete, press the DONE button.
STEP 15|
The main display screen will then appear, On / Off or Standby. Press the ON button to start the
Rapids™ system in the normal operation.
STEP 16|
Press the STANDBY button to place the unit in the standby mode.
STEP 17|
The standby mode will not have any components energized in this sequence, other than the
microprocessor.
Page 45
LabStrong™ Cop.