                                                                                                                        H50/150 Smart – User Manual 

15 

 

connections. 
Connect the ball valve to “DRAIN’’ position of the chiller and turn ball valve off.   
Never connect the fitting to the tap water supply or any water pressure source. 
It is important to keep the distance between the unit and the instrument to be cooled as short 
as possible. Tubing should be straight and without bends. If diameter reductions must be made, 
they should be made on the inlet and outlet of the instrument to be cooled, not on the chiller. 
 
Smart H150 
The liquid plumbing connections are located on the rear of the unit and labeled “SUPPLY” and 
“RETURN”. The connections are 1/2-inch Female Pipe Thread. Units with 1/2-inch fittings are 
supplied with 1/2-inch barbed adapters. 
Remove  the  plastic  protective  caps  from  both  plumbing  connections.  Install  the  barbed 
adapters to these connections. 
Connect  the  fitting  “SUPPLY”  to  the  hose  feeding  the  inlet  of  the  instrument  to  be  cooled. 
Connect the fitting “RETURN” to the hose from the outlet of the instrument to be cooled. Clamp 
all connections. Connect the ball valve to “DRAIN’’ position of the chiller and turn ball valve off.   
Never connect the fitting to the tap water supply or any water pressure source. 
 
It is important to keep the distance between the unit and the instrument to be cooled as short 
as possible. Tubing should be straight and without bends. If diameter reductions must be made, 
they should be made on the inlet and outlet of the instrument to be cooled, not on the chiller. 
 
When fittings have to be changed or the chiller is not used for a long period, be sure to drain all 
liquid out of the unit. Shut down the unit at first, then put a cup on the ground and disconnect 
the fitting from the instrument cooled then let the fluid in the tank flow out into the cup and 
disconnect the fitting on the chiller. Before positioning the unit in the storage be sure the drain 
tap is closed.   
 
3.4 FLUIDS 
Never  use  flammable  or  corrosive  fluids  with  this  unit.  Do  not  use  automotive  antifreeze. 
Commercial  antifreeze  contains  silicates  that  can  damage  the  pump  seals.  The  use  of 
automotive antifreeze will void the warranty. 
 
Fluids should be pure and contain no impurities such as grains. Otherwise, the impurities may 
damage the pump. The use of unpurged fluids will void the warranty. Fluids should be replaced 
every 6 months. 
Whenever fluid is replaced, please kindly add water cleanser into fluid to keep cleaning. For 
H50-500/Smart H150-1000N/Smart H150-1000NLT/Smart H150-2100NS/ 
Smart H150-2100NSLT/Smart H150-3000NS, the volume is 2~3 drops. 
For H150-5000/H150-7000, the volume is 5~6 drops. 
 

Note: please use mixture liquids of water 50% and glycol 50% when setpoint the is lower 
than 5

. Malfunctions caused by using incorrect cooling fluids will void the warranty. 

Summary of Contents for H50 Series

Page 1: ...H50 150 SMART SERIES WATER CHILLERS USER MANUAL...

Page 2: ...invite you to read carefully this user manual and to keep it close to the instrument for convenient and fast consulting For any possible clarification or any request for assistance please contact eith...

Page 3: ...MBING REQUIREMENTS 14 3 4 FLUIDS 15 4 OPERATION PROCEDURE 16 4 1 LCD CONTROLLER 16 4 2 TOUCH CONTROLLER 17 4 2 1 MAIN INTERFACE 17 4 2 2 MAIN MENU 18 4 2 3 ALARMS 20 4 2 4 START UP SHUT DOWN 22 4 2 5...

Page 4: ...tion contained in this document may be the object of patent applications by LABTECH The possession of this document does not confer any license rights in and to such patents The following names are LA...

Page 5: ...following keys FRONT VIEW The front panel consists of the following parts Actual Temperature Alarm Cooling Indication Working State Indication Setpoint Temperature Value Increase Backlight On Off Val...

Page 6: ...lating pump a reservoir and a microprocessor temperature controller Technical specification Model H50 500 115V H50 500 DF Temp control range 5 35 Temp control mode PID Cooling mode Compressor cooling...

Page 7: ...H50 150 Smart User Manual 7 SMART H150 CONTROL PANEL Touch controller H150 Series FRONT VIEW Temperature Controller Liquid Level Indicator...

Page 8: ...H50 150 Smart User Manual 8 REAR VIEW Liquid Supply Connection Liquid Return Connection Pressure Regulation Valve Liquid Drain Electrical Socket Fuse holder...

Page 9: ...n Reservoir volume 2 2L 2 2L 3 5L 3 5L 3 5L 3 5L Recirculating pump PR4 PR4 PR4 PRG8 PRG8 PRG8 Weight 56Kg 56Kg 72Kg 80Kg 80Kg 85Kg Dimension LxWxHmm 560x360x590 650x385x625 740x460x700 NOTE the value...

Page 10: ...H50 150 Smart User Manual 10 MAGNETIC PRG PR SERIES PUMP CAPACITY...

Page 11: ...circuit parts in order to avoid a short circuit or open circuit Fire safety Numerous reagents are flammable and explosive When the solvent vapor concentration reaches a certain level it would be flam...

Page 12: ...the manufacturer s warranty Transport the unit with care Inclination angle must be less than 60 degrees otherwise the refrigeration system could be damaged Sudden jolts or drops can damage the refrige...

Page 13: ...ront of the unit and discharged through the rear and side panels The unit must be properly positioned so that the intake and discharge are not impeded A minimum clearance of 1 meter nearly 3 feet on a...

Page 14: ...PPLY and RETURN Remove the plastic protective caps from both plumbing connections Install the barbed adapters to these connections Connect the fitting SUPPLY to the hose feeding the inlet of your appl...

Page 15: ...oled as short as possible Tubing should be straight and without bends If diameter reductions must be made they should be made on the inlet and outlet of the instrument to be cooled not on the chiller...

Page 16: ...than the setpoint temperature The refrigeration system will stop automatically Alarm code H A will be displayed and the buzzer alarm sound when the temperature is overshoot and 5 or more higher than t...

Page 17: ...hem e Running Alarm status shows the unit status between Running and Alarm Available alarms Temperature Over range Liquid Level Gas Pressure Liquid Pressure Flow Rate Self Testing High Temperature Low...

Page 18: ...confirm the new value touch OK button c Parameter Range parameter setting range d Numeric Keypad keypad can only be used when the value area is activated yellow For negative values touch the button in...

Page 19: ...alue Alarm is selected 20 0 HTL C HDev High deviation value of temperature Active when the Temperature deviation value Alarm is selected 0 0 35 0 C LDev Low deviation value of temperature Active when...

Page 20: ...e factory setting Any malfunction due to improper setting is not covered by the warranty 4 2 3 ALARMS a TEMPERATURE ALARM Temperature Over range Alarm alarm code is shown and the buzzer alarm sounds w...

Page 21: ...e of temperature Low deviation value LDev the buzzer alarm sounds and a warning alarm is shown to indicate over cooling HDev and LDev can be set in the PARAMSET TAlarm window The controller automatica...

Page 22: ...e than 5 seconds When the Water Flow Rate Analogy Amount alarm is activated the touch controller shows flow rate abnormal alarm and the buzzer sounds if the flow sensor detects a flow rate higher than...

Page 23: ...the circular regulation valve opening on the rear of the unit Turn the threaded stem fully counterclockwise If the unit is not connected to the instrument to be cooled install a loop of hose equipped...

Page 24: ...he cup and disconnect the fitting on the chiller Before positioning the unit in the storage be sure the drain tap is closed 5 2 CONDENSER CLEANING For proper operation the unit needs to pull a substan...

Page 25: ...heck the pressure gauge adjust the relief valve as necessary 6 3 INADEQUATE TEMPERATURE CONTROL Verify the setpoint If the temperature continues to rise make sure the instrument to be cooled heat load...

Page 26: ...H50 150 Smart User Manual 26 7 FLOW CHART...

Page 27: ...inators who are ready to quickly assist customers with answers and solutions to service needs and application questions For any possible clarification or any request for assistance please contact eith...

Reviews: