6 OPERATION
94
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
6.9.5.3.
Analog Outputs
Configure the two Analog Outputs as needed per application. These Analog Outputs
have the same options listed in the Local I/O section below.
Analog Output Signal
Select between either Voltage or Current Analog outputs.
•
Voltage Conditions: 0-10 VDC, max load 1000 ohm
•
Current Conditions: 4-20 mA, max load 500 ohms
Local I/O
Three analog outputs can be configured on the Local I/O connector per the following
options:
•
Leak Rate Mantissa
•
Leak Rate Exponent
•
Leak Rate Logarithmic
•
Leak Rate Compound
•
Inlet Pressure
Configure the analog outputs referencing the following information. For the Exponent
and Logarithmic selections, the user must also enter the low decade information.
Mantissa
The "Mantissa" output corresponds with the leak rate mantissa:
Formula
U = Voltage measured (V) on analog output
Mantissa = U
Examples
•
U = 3.5 V -> Mantissa = 3.5