5 INSTALLATION
38
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
Accessory Features
Two accessory ports allow the following functionality to be added to the TITAN
VERSA:
•
High-Flow Evac valve
•
High-Flow Vent valve
•
Pass / Fail light
24 VDC power is provided (fused at 2.0 amps
—
see fuse F1) to all accessory functions.
Table 17 below identifies the function of each pin on accessory connectors.
Accessory Pin Functions
PIN
Function
1 - BRN
Normally closed connection (24 VDC in off position)
3
–
BLU
0 VDC
4 - BLK
Normally open connection
Figure 6:
Accessory Pin Locations
Table 18 below contains a list of current accessories and how to connect them.
After connecting an accessory, go to Menu > System > Accessories to configure the
accessory.
Accessory List, Connections, and Reference
Accessory
Connection
Reference
High-Flow Evac
Always Accs. 1 (internally)
SMT-07-1046
High-Flow Vent
Accs. 1 or Accs. 2 (internally)
Pass / Fail Light
Accs. 1 and Accs. 2 (externally)
This section
The pass / fail light assembly illuminates a pass or fail depending the final leak rate
value vs. the reject limit.