6 OPERATION
82
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
Calibration Method
Select option of either Internal (Vacuum, Mass 4 only) or External calibration per Table 33.
Leak Rate
Enter the leak standard leak rate.
Units
Select the leak standard units.
Date
Enter the month and year of the leak standard calibration in the MMYYYY format.
Cal Temp
Enter the leak standard calibration temperature in Celsius.
Temp Coefficient
Enter the temperature coefficient in 0.1 % precision.
Depletion Rate
Enter the depletion rate. Round to nearest whole percentage number.
Current Temperature
Enter the estimated current leak standard temperature in degrees Celsius (only for
External leaks).
Calibration Start
Cal Start Method
There are three options for calibration start.
•
Cal at Start (not recommended but available)
•
Cal Check at Start (This runs both a cal check and an autocalibration)